Battery testing equipment measures cell performance by quantifying the electrochemical and mechanical responses of a cell under controlled conditions. Core parameters include capacity, voltage, coulombic efficiency, internal resistance, and electrode expansion, each revealing a different aspect of how a cell stores and releases energy. The sections below address the most common technical questions researchers encounter when setting up or interpreting battery test measurements.
What parameters does battery testing equipment actually measure?
Battery testing equipment measures voltage, current, capacity, coulombic efficiency, and internal resistance as its primary electrochemical parameters. These metrics, recorded across charge and discharge cycles at defined C-rates, form the foundation of any quantitative assessment of cell performance. Secondary parameters such as electrode thickness change, gas evolution, and temperature may also be captured depending on the test cell configuration.
In practice, a well-configured battery test system logs the following during cycling:
- Specific capacity (mAh/g or mAh/cm²): the charge stored per unit mass or area of active material, used to evaluate electrode performance independently of cell geometry
- Coulombic efficiency: the ratio of charge extracted on discharge to charge input on charge, expressed as a percentage, a sensitive indicator of irreversible reactions including Solid Electrolyte Interphase (SEI) formation on the anode
- Voltage profiles: the cell potential as a function of state of charge, revealing phase transitions and overpotential contributions
- Internal resistance: derived from voltage response to current steps or from electrochemical impedance spectroscopy (EIS), indicating ohmic and kinetic losses
- Rate capability: capacity retention across a range of C-rates, reflecting how well a cell sustains performance under higher current demands
The choice of which parameters to prioritise depends on the research question. For electrode material development, specific capacity and coulombic efficiency are typically the primary outputs. For cell-level degradation studies, resistance evolution and capacity fade over extended cycling are more informative.
How does electrochemical impedance spectroscopy reveal cell health?
Electrochemical impedance spectroscopy (EIS) reveals cell health by measuring the frequency-dependent impedance of a cell, separating contributions from ohmic resistance, charge-transfer kinetics, and diffusion processes. Each component of the impedance spectrum corresponds to a distinct physical or chemical process, allowing researchers to detect degradation mechanisms that are invisible in standard voltage-capacity measurements.
In a typical EIS measurement, a small sinusoidal perturbation is applied across a range of frequencies, often from a few millihertz to several hundred kilohertz. The resulting Nyquist or Bode plot can be fitted to an equivalent circuit model to extract:
- Ohmic resistance (Rs): electrolyte resistance and contact resistance at current collectors
- SEI layer resistance: the impedance contribution of the Solid Electrolyte Interphase, which typically grows with cycling and contributes to capacity fade
- Charge-transfer resistance (Rct): the kinetic barrier to lithium-ion intercalation at the electrode surface, sensitive to temperature and state of charge
- Warburg impedance: solid-state diffusion of lithium ions within the electrode material, visible at low frequencies
EIS battery testing is particularly valuable for tracking cell ageing non-destructively. An increase in Rct over cycles, for example, often signals electrolyte decomposition or active material isolation, whereas a growing SEI contribution points to continued reductive side reactions at the anode. Instruments capable of both galvanostatic cycling and EIS, such as those with integrated potentiostat/galvanostat (PStat/GStat) functionality, allow researchers to perform impedance measurements at defined states of charge without removing the cell from the test fixture.
What is the difference between galvanostatic and potentiostatic testing?
Galvanostatic testing controls the current applied to a cell and records the resulting voltage response, whilst potentiostatic testing controls the voltage and records the resulting current. The choice between the two modes determines what aspect of cell behaviour is being characterised and which electrochemical processes are being isolated.
Galvanostatic mode
Galvanostatic cycling is the standard method for measuring specific capacity and coulombic efficiency. A defined current, expressed as a C-rate relative to the electrode’s theoretical capacity, is applied during charge and discharge. The voltage profile that results reflects the thermodynamic and kinetic properties of the electrode material, including plateau regions associated with phase transitions and overpotential increases under higher current loads. The galvanostatic intermittent titration technique (GITT) extends this approach by alternating current pulses with rest periods to separate equilibrium potential from kinetic contributions.
Potentiostatic mode
Potentiostatic control is applied when the voltage must be held precisely at a defined value, for example, at the end of a constant-current charge step to ensure full lithiation, or during cyclic voltammetry (CV) to map redox processes as a function of potential. In CV, the potential is swept at a defined scan rate and the resulting current response reveals oxidation and reduction peaks corresponding to lithium insertion and extraction reactions. Potentiostatic EIS is a related technique in which the impedance is measured under voltage control at a fixed potential.
Most modern battery test systems support both modes within the same instrument, allowing researchers to combine galvanostatic cycling with periodic EIS or CV measurements within a single automated protocol.
How do dilatometers measure electrode expansion during cycling?
Electrochemical dilatometers measure electrode expansion during cycling by continuously tracking the displacement of a plunger resting on the electrode stack as the cell charges and discharges. The thickness change of the electrode, which results from lithium-ion intercalation and deintercalation, is recorded as a function of time, voltage, or capacity with sub-micrometre or even nanometre-level resolution.
Electrode expansion is a direct consequence of the structural changes that occur when lithium ions are inserted into an active material. Graphite, for instance, expands by approximately 10% upon full lithiation, whilst silicon-based anodes can expand by several hundred per cent. Quantifying this expansion is essential for:
- Assessing mechanical stress and its contribution to capacity fade
- Validating computational models of electrode mechanics
- Screening new active materials for volumetric stability
- Detecting gas evolution, which produces an additional force contribution distinguishable from mechanical swelling when a separate gas pressure sensor is included
High-resolution dilatometry requires careful control of the applied stack pressure, as uncontrolled pressure changes during cycling introduce artefacts into the displacement signal. Instruments designed specifically for this measurement, such as the ECD-4-nano, which achieves a displacement resolution of better than 5 nm, use integrated force sensors to separate genuine electrode expansion from compliance effects in the measurement system.
What role does temperature control play in accurate battery testing?
Temperature control is essential for accurate battery testing because virtually every electrochemical parameter, including capacity, internal resistance, coulombic efficiency, and reaction kinetics, is temperature-dependent. Without stable, well-defined temperature conditions, data collected across different sessions or laboratories cannot be meaningfully compared, and degradation mechanisms driven by thermal effects may be misattributed to material properties.
The influence of temperature on cell performance operates through several mechanisms:
- Ionic conductivity: electrolyte conductivity decreases at lower temperatures, increasing ohmic resistance and reducing rate capability
- Charge-transfer kinetics: the rate of lithium-ion intercalation follows Arrhenius behaviour, meaning even modest temperature fluctuations affect Rct measurably
- SEI stability: elevated temperatures accelerate SEI growth and electrolyte decomposition, whilst sub-ambient temperatures can cause lithium plating on graphite anodes
- Mechanical behaviour: binder and separator properties change with temperature, affecting contact resistance and compression behaviour within the cell stack
For reproducible battery performance metrics, the test cell and its environment should be held at a defined temperature throughout the measurement. Integrating the cell chamber directly with the test instrument, rather than relying on an external oven, minimises thermal gradients and reduces the risk of temperature drift between the measurement point and the cell itself.
How do you ensure reproducible data from battery test cells?
Reproducible data from battery test cells requires standardised cell assembly, controlled stack pressure, well-defined temperature conditions, and consistent electrode preparation. Each of these variables, if left uncontrolled, introduces scatter into the measured parameters that can obscure genuine material differences or mask degradation trends.
Several assembly-related factors are particularly significant:
- Stack pressure: uncontrolled or inconsistent compression of the electrode stack changes contact resistance, ionic transport through the separator, and the mechanical constraint on electrode expansion, all of which affect measured capacity and impedance
- Moisture contamination: residual moisture in cell components reacts with lithium-containing materials and electrolyte, generating irreversible capacity loss that is easily confused with material-intrinsic behaviour. Housing materials that absorb moisture, such as PEEK, require extended vacuum drying, whilst alternatives such as PPS reduce this risk
- Electrode homogeneity: uneven compression of the electrode disc leads to non-uniform current distribution, broadening voltage plateaus and reducing apparent capacity
- Plunger integrity: current collector surfaces that embed particles over repeated use gradually alter the cell geometry and contact resistance. Tungsten carbide plungers resist this degradation and maintain consistent geometry across many measurement cycles
Beyond hardware, protocol consistency matters equally. Defining rest periods before measurement, specifying the state of charge at which EIS is performed, and recording the exact C-rate and cut-off voltages used are all necessary for data that can be reproduced within a lab and compared with published results.
How EL-Cell GmbH supports accurate battery performance measurement
EL-Cell GmbH designs and manufactures battery testing equipment specifically for the challenges described above. Our product range addresses the most common sources of measurement error in lab-scale electrochemical testing, from cell assembly failures to uncontrolled stack pressure and moisture ingress.
Key capabilities we offer include:
- Integrated force measurement: PAT-Cell-Force and PAT-Cell-Solid include a built-in force sensor, allowing stack pressure to be monitored continuously throughout cycling rather than only at the point of assembly
- Homogeneous electrode compression: the PAT-Solid-Core insert uses guided plane-parallel tungsten carbide plungers and a dedicated pressing tool to ensure uniform compression across the electrode disc
- Low-moisture cell housings: aluminium seals, glass-metal feedthroughs, and PPS plastic construction reduce moisture contamination compared to O-ring-sealed PEEK housings
- High-resolution dilatometry: the ECD-4-nano measures electrode thickness changes with a resolution of better than 5 nm, suitable for characterising low-expansion materials and detecting early-stage degradation
- Multi-channel testing with EIS: the PAT-Tester-i-16 integrates a battery tester, temperature-controlled cell chamber, and docking station into a single instrument, supporting up to 16 channels with full PStat/GStat and EIS functionality
If you are setting up a new battery research workflow or looking to improve the reproducibility of your existing measurements, contact our team to discuss which configuration best fits your experimental requirements.